Model Number URF-P7
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2019
Event Type  malfunction  
Manufacturer Narrative
The subject device had not been returned to olympus medical systems corp (omsc). But was returned to olympus (b)(4). The subject device was sent to a third party laboratory for additional microbiological testing. In the additional test, the testing indicated no microbial growth for the subject device. The physical findings of the subject device were not reported. Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined at this time. If additional information is received, this report will be supplemented.
 
Event Description
Olympus was informed that the subject device tested positive for staphylococcus non aureus (7 cfu/endoscope) during routine surveillance culturing test by the facility. The portion of the subject device, where the microbes were detected, was not reported. The subject device had been brushed manually using a non-olympus cleaning brush (life partner, model: nova clean, ref: cj-kedb-120-02) and disinfected manually with peracetic acid (anioxyde 1000). There was no report of patient infection associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO FIBERSCOPE
Type of DeviceURETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key8774775
MDR Text Key150890379
Report Number8010047-2019-02477
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation
Type of Report Initial
Report Date 07/10/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-P7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/12/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No