• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number BELLAVISTA 1000
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem Injury (2348)
Event Date 12/19/2018
Event Type  Injury  
Manufacturer Narrative
Any additional information received from the customer will be included in a follow-up report.Latest log files are not provided and device has not been returned to the manufacturer.However, per the customer, the unit is working well after replacement of ssd and power supply.Several attempts to contact the customer for additional information such as exact date of the incident and injury leading to the surgery during the power supply failure were made.
 
Event Description
The customer reported that technical failure (tf) 210 was active while using the bellavista 1000.It was also reported the 24 volt is not available at the power supply outlet.The issue occurred during patient use, and the patient was removed from the ventilator and changed with another one.The customer reported patient injury, and surgery was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs, 9470
SZ  9470
Manufacturer (Section G)
TECHNOCOM SYSTEMS SDN BHD
plo 1 jalan firma 1
kawasan perindustrian tebrau 1
johor bahru, 81100
MY   81100
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key8778305
MDR Text Key150692797
Report Number3004553423-2019-00033
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000
Device Catalogue Number301.100.000
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/19/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-