Model Number 60-02-60 |
Device Problem
Decreased Pump Speed (1500)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/13/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative investigated the unit.He duplicated the failure but he could not identify the cause.Further checks will be conducted in order to determine the cause of the failure.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) did not rotated at the expected speed during priming.After a while, it operated properly again.However, the user decided to replace the pump with another one as precaution.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the reported failure could be solved by rebooting the device and did not recur.The root cause could not be determined.
|
|