Model Number 60-02-60
Device Problem Decreased Pump Speed (1500)
Patient Problem No Patient Involvement (2645)
Event Date 06/13/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative investigated the unit.He duplicated the failure but he could not identify the cause.Further checks will be conducted in order to determine the cause of the failure.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) did not rotated at the expected speed during priming.After a while, it operated properly again.However, the user decided to replace the pump with another one as precaution.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the reported failure could be solved by rebooting the device and did not recur.The root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8780518
MDR Text Key150713688
Report Number9611109-2019-00530
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817901006
UDI-Public010403381790100611170727
Combination Product (y/n)N
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Was the Report Sent to FDA? No
Date Manufacturer Received09/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-