It was reported that the procedure was to treat a moderately calcified lesion in the superficial femoral artery (sfa).
The 4.
0x200mm armada 18 balloon catheter was advanced to the target lesion and inflated to nominal pressure when it was noted the proximal balloon marker was located inside the balloon working length, not on the shoulder where it should be.
Although the balloon markers were slightly off, the balloon worked as intended.
The same balloon was used to successfully complete the procedure.
There was no clinically significant delay in the procedure and no adverse patient effects.
No additional information was provided.
|