Model Number 1009-9050-000 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).For 5 of the 5 reported events, a ge healthcare service representative performed a checkout of the system and confirmed the reported events.To resolve the reported events, the flow sensor was replaced.
|
|
Event Description
|
This report summarizes 5 malfunction events.A review of the events indicated that model 1009-9050-000 anesthesia gas machine experienced flow sensor malfunction that cause a loss of mechanical ventilation.These reports were received from various sources.Of the 5 events, 4 did not involve patients, and 1 did involve a patient.There was no patient information available.
|
|