MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Catalog Number 2000-03
Device Problems Insufficient Cooling (1130); Mechanical Problem (1384); Defective Component (2292)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed. The device was not returned.
 
Event Description
It was reported that the arctic sun device did not cool the water during therapy. Therapy was unable to be completed due to the mechanical failure, which was determined to be caused by a chiller malfunction. Per additional information received via email on 30jun2019, the device was repaired and the issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key8788990
MDR Text Key151114980
Report Number1018233-2019-03892
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/20/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number2000-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-