BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7010
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/13/2019
Event Type  malfunction  
Manufacturer Narrative
Returned product consisted of a quantum maverick balloon catheter. The shaft, hypotube, tip and balloon were microscopically and visually examined. There was contrast in the inflation lumen. The balloon was tightly folded. There were numerous hypotube kinks throughout the device. There was a complete hypotube separation 50cm distal of the strain relief. The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation. The tip was damaged. Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 24jun2019. It was reported that shaft kink occurred. The target lesion was located in the left coronary artery. A 2. 5mm x 12mm quantum maverick balloon catheter was advanced for dilatation. However, it was noted that the delivery shaft was kinked. The procedure was completed with another of the same device. No patient complications were reported that the patient's status was stable. However, returned device analysis revealed separated patient's.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUANTUM MAVERICK
Type of DeviceCATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key8789957
MDR Text Key151012678
Report Number2134265-2019-08196
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 07/15/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/22/2022
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0023230378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-