The physician was called with patient tachypnea and oxygen desaturation, respiratory rate 40.The ventilator was set for a rate of 12 and the patient was not breathing on own, but the vent was showing that the patient was over breathing on the graph.Sedation was increased - propofol, versed, dilaudid - with no relief.Patient sedated to 83%.Continued propofol, versed, dilaudid and nimbex x1 dose given.A vent issue was noted so the ventilator was replaced by respiratory therapy.After all of the above, the respiratory rate returned to normal.It was confirmed it was a vent issue and not a sedation/patient issue.Follow up: respiratory therapy staff examined the enve ventilator and determined that excessive moisture in the pressure line was causing the autocycling.The patient circuit was replaced, and the ventilator ran for 8 hours with no problems.Vent next brought to biomedical engineering and checked history and nothing relevant was noted.Ran unit.Determined with respiratory therapy manager that unit was ok to return to service.
|