Model Number N/A
Device Problem Material Separation (1562)
Patient Problem No Code Available (3191)
Event Date 07/04/2019
Event Type  Injury  
Manufacturer Narrative
Pma/510(k) #: preamendment.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It was reported an ultrathane mac-loc locking loop biliary drainage catheter was placed in an unknown patient as a chest tube.It was reported that the drain came apart after it was placed in the patient.The patient is scheduled to have a second procedure to replace the device.No other adverse effects were reported for this incident.
 
Manufacturer Narrative
B5- additional information received 22jul2019 provided more event details.The device was place with via "yueh centesis needle, rosen wire, dilator, then over the wire." no unusual resistance was felt during placement.Upon first inspection, the fitting was securely attached, but the separation at the tube connection to mac-loc was found the next morning.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Investigation ¿ evaluation.A review of the complaint history, drawing, documentation, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.The proper procedures are in place to identify and prevent this failure mode prior to device distribution.The risk specifications covering mac-loc drainage catheters include both hub separation and leakage as a potential failure mode.The identified risk controls include manufacturing quality control checks and process validation.The technical files covering mac-loc drainage catheters indicate that the risks associated with these devices are acceptable when weighed against the benefits.The lot number of the device is not known; accordingly, a review of the device history record could not be conducted.Based on the information provided, no returned product and the results of the investigation, a definitive cause could not be established appropriate measures have been taken to address this failure mode.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Additional information: b5 - previous mdr submission stated "the patient is scheduled to have a second procedure to replace the device." additional information received 11jul2019 reported the patient had a successful procedure to replace the faulty drain.The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER
Type of Device
GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8793378
MDR Text Key151124699
Report Number1820334-2019-01694
Device Sequence Number1
Product Code GCA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULT12.0-38-40-P-32S-CLB-RH
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-