Model Number U228
Device Problems Over-Sensing (1438); Off-Label Use (1494); Device Sensing Problem (2917)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/12/2019
Event Type  Injury  
Manufacturer Narrative
This case is complete.Upon further information this investigation will be updated.
 
Event Description
It was reported that noise resulting in oversensing was observed on the right ventricular channel as this device was implanted without a right ventricular lead due to a biological valve implant and the right ventricular port of the device was plugged.Boston scientific technical services (ts) discussed that the noise was due to the auto lead detect which resulted in oversensing with left ventricular inhibition seen for less than one second.A revision took place the right atrial lead was connected to right ventricular port to deactivate auto lead detect and the physician connected the is-1 plug again, after which no noise was seen on the right ventricular channel and no left ventricular pace inhibition.All measurements were normal.The device remained implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST X4 CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CLONMEL LIMITED
cashel road
,
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8793967
MDR Text Key151150349
Report Number2124215-2019-13657
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559471
UDI-Public00802526559471
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/15/2020
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number736605
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-