MAUDE Adverse Event Report: MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Catalog Number 209999
Device Problems Electrical /Electronic Property Problem (1198); Device Difficult to Setup or Prepare (1487)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/03/2019
Event Type  malfunction  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical. A supplemental report will be submitted when additional information becomes available.
 
Event Description
Case number: (b)(4), mps (b)(6) reported hybrid cable loose and difficulty getting fisso arm to tighten. Update: "problem noticed: loose cables were noticed intra-op prior to making cuts when there was a software error that came up. Case type: pka. Surgical delay: 20 minutes. Update: was the patient under anesthesia? as per mps: "yes the patient was under anesthesia". Case cancelled: no. Case completed successfully: yes. Case completed manually: no, i was able to re-tighten cables as well as troubleshoot the software issue that the system was giving me. We were able to complete the case robotically. Investigation results: yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name3.0 RIO® ROBOTIC ARM - MICS
Type of DeviceSTEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
bethany hinson
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key8795406
MDR Text Key155873631
Report Number3005985723-2019-00515
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number209999
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/30/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 07/16/2019 Patient Sequence Number: 1
-
-