Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was functionally tested with a.014 guidewire.Guidewire was unable to be inserted thru the tip due to tip damage.Guidewire encountered resistance at tip when advanced thru exit notch.Inspection of the remainder of the device presented no other damage or irregularities.
|