• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TEAR DROP GUIDE WIRE 3.0 MM DIAMETER 100 CM LENGTH; ROD, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TEAR DROP GUIDE WIRE 3.0 MM DIAMETER 100 CM LENGTH; ROD, FIXATION Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Date 09/07/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned products identified the presence of a crease and/or fold in the seal of all of the returned products.These creases and/or folds are unacceptable.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing deficiency.A corrective action has been initiated to address the manufacturing deficiency.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the seal was peeled off and there was a crease in the sealing area.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEAR DROP GUIDE WIRE 3.0 MM DIAMETER 100 CM LENGTH
Type of Device
ROD, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8801107
MDR Text Key151445419
Report Number0001822565-2019-03056
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K101622
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model NumberN/A
Device Catalogue Number47249009700
Device Lot Number64101468
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Was the Report Sent to FDA? No
Date Manufacturer Received07/02/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0542-2019
Patient Sequence Number1
-
-