The subject device was returned with two clips.The coil sheath of the subject device (clip fixing device) was bent at 460 mm from the distal end.We checked the distal end of the subject device, there was no abnormality such as the deformation.We felt stuck when we pulled the handle of the subject device.The first clip was closed.The clip connector of the first clip was broken.Something biological adhered to the first clip.The second clip was closed.The clip connector of the second clip was not broken.Something biological did not adhere to the second clip.When we put an unused clip on the subject device and we tried clipping, it could be done.However, we needed more force to operate handle for clipping.The manufacturing record was reviewed and found no irregularities.There was a possibility that the clip could not be released because the operation of the slider became heavy and difficult to be pulled out all the way to the end due to such as the bent of the coil sheath.The above device handling has warned in the instruction manual as follows.Do not withdraw the instrument if clipping is not completely finished (when the slider is not pulled out all the way in the proximal direction).Doing so may tear tissue inside the body cavity, resulting in patient injury, such as perforation, bleeding, or mucous membrane damage.
|