Qn# (b)(4).The customer returned one 4-lumen catheter, 60ml syringe and tubing for evaluation.The catheter body was observed to be cut short.Visual examination of the catheter did not reveal any defects or anomalies.The overall length of the catheter body measured 91mm which is not within specification.The catheter body was observed to be cut, most likely during the procedure.The outer diameter of the proximal line and distal line were measured and both were found to be within specification.The returned catheter was tested for liquid leakage.The catheter was attached to the leak tester and the distal end was occluded, and the leak tester was turned to 300 kpa for 30 seconds.No leaks or air were detected from any region of the catheter, including the extension lines; therefore, the sample passed functional testing.All 4 extension lines were tested.The ifu provided with this kit warns the user "do not cut catheter to alter length".The reported complaint of the catheter issue could not be confirmed through examination and functional testing of the returned sample.The returned catheter did not leak during functional testing and met all relevant dimensional and visual requirements.No problem was found with the returned catheter.Teleflex will continue to monitor and trend for complaints of this nature.
|