(b)(4).Concomitant medical products: 010000740 lot 6500012 g7 neutral arcomxl lnr 36mm e, (01)00880304524972, 010000740 lot 6499982 g7 neutral arcomxl lnr 36mm e, (01)00880304524972, 110024463 lot 857940 g7 dual mobility liner 42mm e, (01) 00880304703513, 010000663 lot 6549354 g7 pps ltd acet shell 52e, (01) 00880304524224.Multiple mdrs were filed for this event.Please see associated: 0001825034 - 2019 - 03060, 0001825034 - 2019 - 03061, 0001825034 - 2019 - 03062, 0001825034 - 2019 - 03063.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
D4 - udi # - (b)(4).D11 - medical products - unknown screw qty: 2.Visual examination of the returned product identified: one g7 pps ltd acet shell 52e ((b)(4)), two g7 neutral arcomxl lnr 36mm e ((b)(4)) ((b)(4)), one g7 dual mobility liner 42mm e ((b)(4)), and two unknown legacy zimmer screws were returned and evaluated.Upon visual inspection of the two arcomxl lnrs ((b)(4) and (b)(4)) the locking feature shows impact markings.The dual mobility liner shows damage to the post.There was no damage to the locking feature of the shell.Review of the device history record identified no deviations or anomalies during manufacturing.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|