MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number JC-05400-B
Device Problem Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that a triple incident occurred on (b)(6) 2019 in the maternity room.Three different doctors faced the same issue with devices from the same lot number.They had difficulties inserting the catheter in the tuohy needle because there was a blockage at the level of the distal bevel.
 
Event Description
It was reported that a triple incident occurred on (b)(6) 2019 in the maternity room.Three different doctors faced the same issue with devices from the same lot number.They had difficulties inserting the catheter in the tuohy needle because there was a blockage at the level of the distal bevel.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter and needle with no relevant findings.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter and needle with no relevant findings.The potential cause of catheter not threading could not be determined based upon the information provided and without a sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8811969
MDR Text Key151738279
Report Number3006425876-2019-00517
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/07/2021
Device Catalogue NumberJC-05400-B
Device Lot Number71F19D1772
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-