MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number RTLR180343
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2019
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.There were visual indications of dried fluid on the top cover within the cassette compartment.There were no visual indications of particulates within the cassette area.There were no burrs or sharp edges in cassette area that may have punctured a cassette membrane.Upon power up, the cycler touch screen test failed.When powering on the cycler, the ok, stop and up/down arrow push buttons illuminated; however, the front panel touch screen remained blank.A burning smell did not occur.It was identified that the cause for the blank screen was due to an internal short present transformer (t1) on the inverter board.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the display became fully operational.Visual evidence of dried fluid was found on the bottom cover adjacent to the front panel assembly during the internal inspection.The cause of the observed dried fluid could not be determined.An investigation of the cycler mushroom heads verified that the surface conditions and alignments were within specification.The cycler tested positive for glucose.A device history record (dhr) review was performed and found a prior occurrence of blank screen.A failure analysis problem report indicated that the failure was due to a defective inverter board.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be an internal short on transformer on the inverter board.The cycler was refurbished following the evaluation.
 
Event Description
It was reported that the screen of a patient¿s liberty select cycler went blank during drain two of their peritoneal dialysis (pd) treatment.The patient stated that they started to smell a burning smell.The power cord was properly connected in both ends and cycler plugged into a power strip that was shared with the modem and a laptop.The outlet was working.The ok and stop keys were on.The cycler was rebooted; however, the screen remained blank.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was not available.Upon follow up, the pdrn confirmed that there were no adverse events or medical intervention required as a result of the reported event.The patient continued treatment on the cycler.Upon physical evaluation of the cycler by the manufacturer, evidence of an internal short on the transformer on the inverter board was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8812722
MDR Text Key151760657
Report Number2937457-2019-02254
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2019
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2019
Device Age MO
Date Manufacturer Received07/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY CYCLER SET 
-
-