CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number RTLR180343 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.There were visual indications of dried fluid on the top cover within the cassette compartment.There were no visual indications of particulates within the cassette area.There were no burrs or sharp edges in cassette area that may have punctured a cassette membrane.Upon power up, the cycler touch screen test failed.When powering on the cycler, the ok, stop and up/down arrow push buttons illuminated; however, the front panel touch screen remained blank.A burning smell did not occur.It was identified that the cause for the blank screen was due to an internal short present transformer (t1) on the inverter board.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the display became fully operational.Visual evidence of dried fluid was found on the bottom cover adjacent to the front panel assembly during the internal inspection.The cause of the observed dried fluid could not be determined.An investigation of the cycler mushroom heads verified that the surface conditions and alignments were within specification.The cycler tested positive for glucose.A device history record (dhr) review was performed and found a prior occurrence of blank screen.A failure analysis problem report indicated that the failure was due to a defective inverter board.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be an internal short on transformer on the inverter board.The cycler was refurbished following the evaluation.
|
|
Event Description
|
It was reported that the screen of a patient¿s liberty select cycler went blank during drain two of their peritoneal dialysis (pd) treatment.The patient stated that they started to smell a burning smell.The power cord was properly connected in both ends and cycler plugged into a power strip that was shared with the modem and a laptop.The outlet was working.The ok and stop keys were on.The cycler was rebooted; however, the screen remained blank.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was not available.Upon follow up, the pdrn confirmed that there were no adverse events or medical intervention required as a result of the reported event.The patient continued treatment on the cycler.Upon physical evaluation of the cycler by the manufacturer, evidence of an internal short on the transformer on the inverter board was identified.
|
|
Search Alerts/Recalls
|
|
|