BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
Model Number 7015 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 24.1cm from the hub.Microscopic examination revealed damage to the tip.There is contrast present in the inflation lumen and balloon.There is blood present in the guidewire lumen and the balloon is loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
|
Event Description
|
Reportable based on device analysis completed on 28jun2019.This device was received with no reported issues.However, upon review of this device, the hypotube was separated.
|
|