MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 130 RAD DRILL GDE DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 130 RAD DRILL GDE DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71674003
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2019
Event Type  malfunction  
Event Description
It was reported that the drill guide would not snap the lag screw trocar in during surgery.No delay reported.Backup was not available so they had to hold drill guide in.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device was manufactured in 2008 and exhibits signs of extensive wear/ usage.The functional evaluation could not confirm the stated failure mode.The device functioned as intended with its mating parts.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
130 RAD DRILL GDE DROP
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8814264
MDR Text Key151850689
Report Number1020279-2019-02717
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560476
UDI-Public03596010560476
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71674003
Device Lot Number08HGT0003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2019
Date Manufacturer Received06/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-