Model Number 16-02-80
Device Problem Blocked Connection (2888)
Patient Problem No Patient Involvement (2645)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a stuck circulator motor.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The affected part has been requested back for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message and gave an alarm, when switching on the unit.This was observed during maintenance.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the replaced component was requested to return for further investigation.Results of the investigation on components revealed that the pump motor shaft was stiff and the motor had a bearing damage.The cause is the rust formed which generates irregularities on the surface of the balls of the bearing.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8821163
MDR Text Key152081413
Report Number9611109-2019-00562
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received09/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-