ABBOTT VASCULAR XIENCE SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number UNK RX SIERRA
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/02/2019
Event Type  malfunction  
Manufacturer Narrative
Date of event: estimated date of event.The customer reported the device was discarded.Investigation is not yet complete.A follow-up will be submitted with all relevant information.
 
Event Description
It was reported in a general comment made by the physician that there is poor visibility of the xience sierra stents under fluoroscopy and non-abbott stents have been favored.There were no reported adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Unique device identifier (udi#): in the absence of a reported part number, the udi cannot be calculated.Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record and complaint history could not be conducted because the part and lot numbers were not provided.The investigation was unable to determine a conclusive cause for the reported difficulty positioning (poor visibility during fluoroscopy); however, factors that may contribute to difficult positioning include, but are not limited to, material issues, incorrect contrast mix and incorrect device preparation.Additionally, it should be noted that this observation was presented during a general conversation and is not related to a specific event and/or procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key8822990
MDR Text Key152084743
Report Number2024168-2019-10308
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK RX SIERRA
Was Device Available for Evaluation? No
Date Manufacturer Received08/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-