• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM G5 SYSTEM; CONTINUOUS GLUCOSE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM G5 SYSTEM; CONTINUOUS GLUCOSE MONITOR Back to Search Results
Model Number 1000096
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Diabetic Ketoacidosis (2364); Lethargy (2560)
Event Date 07/03/2019
Event Type  Injury  
Manufacturer Narrative
The cause of the reported issue was due to the infusion set.Tandem does not manufacture infusion sets.Should new relevant information become available, a supplemental report will be submitted.No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.Device not returned.
 
Event Description
It was reported that an occlusion alarm occurred.Customer's blood glucose (bg) was 403 mg/dl with high ketone levels.A manual insulin injection was administered by customer¿s mother to address elevated bg/ketones.Reportedly, customer was lethargic.Troubleshooting with tandem technical support (cts) determined that the occlusion alarm was due to the infusion set site.Customer did not respond to attempts to obtain additional information and customer could not troubleshoot with cts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM G5 SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
julia kensick
11075 roselle st.
san diego, CA 92121
8583666900
MDR Report Key8832143
MDR Text Key152366597
Report Number3013756811-2019-47443
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00853052007318
UDI-Public00853052007318
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number1000096
Device Catalogue Number1000911
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
INFUSION SET: UNKNOWN
Patient Outcome(s) Other;
Patient Age16 YR
-
-