• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE MONARCH III IOL DELIVERY SYSTEM, INJECTOR; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE MONARCH III IOL DELIVERY SYSTEM, INJECTOR; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number III
Device Problems Failure to Align (2522); Device Displays Incorrect Message (2591); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2019
Event Type  malfunction  
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A complaint history examination indicates there are no additional complaints associated with the lot for the reported issue.Based on the preliminary investigation findings, there has been no change in criticality for this complaint.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A purchasing agent reported that during an intraocular lens (iol) implant procedure, the injector misaligned and scratched the lens.It was reported there was a loading error and there was no patient contact.The procedure was completed the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, INJECTOR
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key8837097
MDR Text Key152524074
Report Number2523835-2019-00338
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIII
Device Catalogue Number8065977773
Device Lot Number711211M
Was Device Available for Evaluation? No
Date Manufacturer Received07/01/2019
Date Device Manufactured10/11/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MONARCH III D CARTRIDGE; SND1T5 11.5D
-
-