Model Number W4TR01
Device Problem Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/17/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the cardiac resynchronization therapy pacemaker (crt-p) exhibited a diagnostic reset. The crt-p remains in use and is scheduled to be interrogated in the clinic by an updated programmer. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed. Analysis of the device memory indicated a software error. Analysis of the device memory indicated an issue with missing/invalid diagnostic data. Analysis of the device memory indicated that the atrial and ventricular rate histogram data was missing/invalid. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePERCEPTA QUAD CRT-P MRI SURESCAN
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8861658
MDR Text Key153294063
Report Number3004209178-2019-14992
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/28/2019
Device Model NumberW4TR01
Device Catalogue NumberW4TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 08/05/2019 Patient Sequence Number: 1
Treatment
459888 LEAD, 5076-52 LEAD, 5076-58 LEAD
-
-