Complaint is confirmed.Received one trs175sb2 in opened original packaging.Lot number was confirmed.Performed a visual inspection of the device, the string was wrapped multiple times around the plunger creating tension.The metal prongs were protruding out of the sides of the retrieval bag.Performed a functional inspection, the white inner plunger was difficult to maneuver in and out.Without disconnecting the string and retrieval bag you could not pull the white plunger out from the purple shaft.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been 9 complaints regarding 9 devices for this device family and failure mode.During the same time frame 385,738 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed, the rate of failure would be.00002 per the instructions for use, the user is advised the following; only physicians trained in the techniques of laparoscopic or minimally invasive surgery and the use of retrieval bags to remove tissue should use the product.Care should be taken when using sharp and electro-surgical devices.This issue will continue to be monitored through the complaint system to assure patient safety.
|