Model Number N/A
Device Problem Fracture (1260)
Patient Problem Infiltration into Tissue (1931)
Event Date 07/14/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a development lab training, a stem was implanted into a stripped femur.While attempting to remove the stem from the inserter handle, the threaded distal tip fractured and remained lodged in the stem.An osteotomy was then performed in order to remove the stem component.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Udi : (b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned device notes threads had fractured inside of the stem which was alos returned.There are impact marks on the shaft towards the rubber handle.Dhr was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to off label usage and failure to follow surgical technique as the implant and instruments are not compatible.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOBLOCK INSERTER HANDLE
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8864251
MDR Text Key153375615
Report Number0001825034-2019-03421
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110024780
Device Lot Number656480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2019
Was the Report Sent to FDA? No
Date Manufacturer Received11/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
11-300814 ¿ ARCOS STEM ¿ 330040; 11-300814 ¿ ARCOS STEM ¿ 330040
Patient Outcome(s) Hospitalization; Required Intervention;
-
-