The device was returned to numed for review.The arrived intact and all components were accounted for.The catheter has traces of blood inside the tubing.A longitudinal balloon burst is confirmed.There are no kinks in the tubing.The report states that the balloon burst at the rbp of 2.0 atm.A dvd was returned with the device and was reviewed.It looks like the balloon performed as it should, but patient anatomy may have contributed to the balloon burst.Two catheters were pulled and tested for rbp.Both of these catheters were the same catalog number as the complaint catheter, but have different lot numbers.A 0.035" guidewire was inserted in each catheter.The catheter balloons were then inflated with an inflation device with pressure gauge up to the rbp of 2.0 atm.No issues were noted.The catheter balloons were then inflated beyond labeled rbp until they burst.One balloon burst at 4.0 atm and one burst at 3.8 atm, which is well above the labeled rated burst pressure of 2.0 atm.Both were clean longitudinal bursts.A component review of the balloon material used on the complaint device was performed.A total of 328 balloons have been made and issued to devices using the same balloon material as the complaint catheter.No other complaints have been received on this balloon material lot number.
|