Model Number MSERIES BI-PHASIC
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.

 
Event Description

Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib pad short" message. Complainant indicated that there was no patient involvement in the reported malfunction.

 
Manufacturer Narrative

The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a fractured solder joint on a capacitor on the high voltage module. The device was recertified and returned to the customer. Analysis for reports of this type has not identified an increase in trend.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameM SERIES
Type of DeviceDEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8870992
MDR Text Key153653574
Report Number1220908-2019-02170
Device Sequence Number1
Product Code DPS
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK990762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/17/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/07/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator OTHER
Device MODEL NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer07/18/2019
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/17/2019
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured09/01/2004
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-