Model Number MSERIES BI-PHASIC |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib pad short" message.
Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a fractured solder joint on a capacitor on the high voltage module.
The device was recertified and returned to the customer.
Analysis for reports of this type has not identified an increase in trend.
|
|
Search Alerts/Recalls
|