|
Model Number 2420-0007 |
Device Problem
Stretched (1601)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.Although requested, no additional information, (including patient information), has been received.
|
|
Event Description
|
It was reported that the nurse was unable to flush the line, and then the pump segment ballooned.
|
|
Manufacturer Narrative
|
The customer¿s report that the pump segment ballooned was confirmed.Visual inspection of the set noted that the silicone segment tubing had a bulge (.8065 inches long,.4980 inches wide), discoloration, and was weakened just below the upper fitment.Clear liquid was observed inside both of the set¿s tubing and drip chambers.No other anomalies were observed.Functional testing resulted in liquid flowing freely with no issues observed.The set was loaded into a lab pump module device.The primary infusion was programmed at a rate of 120ml/h and vtbi 120ml for 1 hour.No alarms or any other issues were noted by the device.No bulge/ balloon was observed in the silicone segment.The set was attached to an air pump and completely submerged in warm water.A balloon appeared.The root cause for the source of the excessive pressure was not determined.The customer¿s report that the nurse was unable to flush was not confirmed.Functional testing was performed and the set was able to flush.
|
|
Event Description
|
It was reported that the nurse was unable to flush and then the pump segment ballooned.
|
|
Search Alerts/Recalls
|
|
|