MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.(b)(6).Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in a (b)(6).The affected part was requested to be returned to livanova (b)(4) for a detailed investigation.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 roller pump displayed an error message.Further details are pending submission.There was no report of patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the reported issue could be confirmed during the investigation at the manufacturer site.The analysis of the pump read-out revealed a deviations with the shaft encoder signal.It was determined that one of the channels of the encoder did not send any signal.The root cause of the reported malfunction could be addressed to a defective shaft encoder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8872664
MDR Text Key174888634
Report Number9611109-2019-00606
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)180920
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-