• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event Description
During an unspecified therapeutic procedure, the subject device was used.In the procedure, an endoscopic image displayed on the monitor was black out.The user replaced the subject device with another device and completed the procedure.There was no patient injury reported.After this event, an olympus service engineer visited the user facility and found the subject device could not be turned up.Therefore, it was determined that the subject device required repair.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.However, according to repair history of the subject device, the power supply unit was replaced since it was found the power supply unit was broken after this event.The device history record was reviewed and found no irregularities.It had already passed about six years since the user used the subject device.Based on the evaluation result so far, omsc surmised that the reported failure phenomenon was attributed to the broken power supply unit.And it is possible that the power supply unit was broken by aging degradation.The clv-s190 instruction manual states the corresponding method when there is an abnormality for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8872933
MDR Text Key204283687
Report Number8010047-2019-02897
Device Sequence Number1
Product Code GCT
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-