Model Number 400CE
Device Problem Insufficient Heating (1287)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/10/2019
Event Type  malfunction  
Manufacturer Narrative
User reported the hemotherm 400ce would not heat over 29c.Complaint #(b)(4)was opened and is being investigated.No adverse event was reported.
 
Event Description
User reported the hemotherm 400ce heater-cooler was heating slowly.The unit would not heat over 29c.No adverse event was reported.
 
Manufacturer Narrative
Device evaluated and determined to have a defective heater.The heater was replaced to correct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM
Type of Device
HEATER-COOLER
Manufacturer (Section D)
GENTHERM MEDICAL, LLC
12011 mosteller
cincinnati OH 45241
Manufacturer (Section G)
GENTHERM MEDICAL, LLC
12011 mosteller
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller
cincinnati, OH 45241
8009897373
MDR Report Key8881214
MDR Text Key194341223
Report Number1516825-2019-00010
Device Sequence Number1
Product Code DWC
UDI-Device Identifier10613031860221
UDI-Public(01)10613031860221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-