RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 10aug2019.The product support engineer (pse) troubleshot this issue with the customer over the phone.The customer reported that he replaced the solenoid 2 and 4 assembly.The customer was advised to replace the solenoid 3 and 4, and to replaced the data acquisition (da) printed circuit board (pcb) if the solenoid did resolve the issue.Multiple attempts were made to contact the customer to gather additional details; however, all attempts were unsuccessful and no additional information was provided.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
 
Event Description
The customer reported that the ventilator generated an error (110a) relevant to proximal pressure sensor autozero failed.The device was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8881552
MDR Text Key154137444
Report Number2031642-2019-05749
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/29/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-