• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - CARTAGO TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CARTAGO TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number 2C4041
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 07/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a y-type tur/bladder irrigation set leaked "from connection of tube" (not further specified).This issue was identified during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: the device was received for evaluation.A visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including clear passage and pressure testing were performed which revealed a leak between the tubing inside the drip housing.The reported problem was verified.The cause of the condition was due to manufacturing issue.There are controls in place for risk reduction in order to prevent and detect leaks.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUR IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8883671
MDR Text Key153984418
Report Number1416980-2019-04373
Device Sequence Number1
Product Code LJH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K960787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4041
Device Lot NumberR18G25028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-