MAUDE Adverse Event Report: B.BRAUN SURGICAL SA DAFILON BLUE 4/0 (1.5) 45CM DS19; OTHER SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA DAFILON BLUE 4/0 (1.5) 45CM DS19; OTHER SUTURE Back to Search Results
Model Number C0932205
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Wound Dehiscence (1154)
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with the dafilon suture after a skin biopsy.One week postoperatively, the wound was still gaping and the physician removed the suture immediately.It was not clear why the dehiscence had occurred.There was no additional information provided.
 
Manufacturer Narrative
D4 expiration date: 07/27/2023.Manufacturing evaluation: samples received: one unopened pouch to analyze the six cases (b)(6)).Analysis and results: there are no previous complaints of this code-batch.There are no units in our stock.We have received one closed sample to analyze the six cases.We have tested the knot pull tensile strength of the sample received and the result fulfils the requirements of the european pharmacopoeia (ep).This value is the usual and the current one for this thread and size.We have also tested the knot security control of the closed sample received and the result is into the current range for this thread and size.Remarks: as stated in the instructions for use (ifu) of the product, "dafilon should be used applying the standard surgical suturing and knotting techniques, taking into account the surgeon's experience with the respective surgical procedure.Care should be taken that the knots are positioned properly and adequate knot security is given".On the other hand, "the following side effects may be associated with the use of this product: tissue injury, transitory local irritation, transient inflammatory foreign body reaction, enhanced bacterial infectivity, wound dehiscence, hypertrophic scar/granuloma formation, track bleeding, increased risk of aneurism and stenosis".Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Final conclusion: according to the results of the sample tested and the batch manufacturing record review, the product complies with our specifications and also fulfill usp/ep requirements.Therefore, we do not see any manufacturing fault or material defect that could have caused the incidence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 4/0 (1.5) 45CM DS19
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key8884182
MDR Text Key153994341
Report Number3003639970-2019-00600
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
PMA/PMN Number
K990090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0932205
Device Catalogue NumberC0932205
Device Lot Number618305
Was Device Available for Evaluation? No
Date Manufacturer Received09/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-