Catalog Number 8065751763 |
Device Problem
Blocked Connection (2888)
|
Patient Problem
Eye Burn (2523)
|
Event Date 07/10/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A nurse reported that the phaco tip blocked during a procedure.The patient experienced a corneal burn.The wound was sutured.Additional information was received.The reporter informed that the handpiece blocked at the commencement of sculpt.The patient experienced a corneal burn with localized corneal opacity.The patient also experienced wound leakage so three sutures were placed to close the wound.The patient developed regular astigmatism form the tight sutures.In the surgeon's opinion, occlusion of the handpiece caused the event.Additional information was requested and received.
|
|
Manufacturer Narrative
|
The phaco handpiece was not returned for evaluation.The phaco handpiece manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.With no additional, related information provided, the customer reported events could not be confirmed.The root cause of the reported event is unable to be determined conclusively.(b)(4).
|
|