The initial reporter stated that are having issues with mismatched patient identifiers on two cobas u 411 analyzers (serial numbers (b)(4)).When scanning the barcode of patient samples, the identifiers printed by the analyzer are different.The issue occurred at least twice.With each analyzer, sample barcodes are scanned, "v" is pressed on the sample entry window of the system software, and then a test strip is placed on the tray.For u 411 serial number (b)(4), the reporter mentioned that they have tried re-setting the sequence number in the system software but the reset does not work.For both systems, the reporter mentioned that when processing a total of 10 patient samples, for example, 10 test strips are inserted into the analyzer, but not all results are printed.Some samples from the work list generated in the system software do not disappear.An error is encountered which states there was measurement inconsistency.The error message instructs the user to confirm the correct work list sequence and try again.
|