H.10: the unit was requested by livanova deutschland for a further and detailed investigation.No errors occurred during the test and no deviations could be detected.The reported malfunction could not be reproduced.Based on the above, the root cause of the reported event could not be determined.Since the error was reproduced during the visit of the field service representative at the customer's site and not during the intensive tests at livanova deutschland, it is reasonable to think that the issue is not device-related, but could be due to the hospital gas supply.According to the instruction for use, a minimum pressure of 2 bar per connection should be observed.Indeed, input pressure too low can cause a discrepancy in the target and actual values, leading to the reported error code.
|