Model Number 3662
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem Inadequate Pain Relief (2388)
Event Date 07/25/2019
Event Type  Injury  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event Description
It was reported that the patient's scs ipg would not communicate with external devices following a fall on (b)(6) 2019. A manufacturer representative met with the patient and confirmed the issue, deeming the device inoperable. The patient may undergo surgical intervention to resolve the issue.
 
Manufacturer Narrative
The patient underwent an unrelated surgery where electro-cautery was used. A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system. The results of the investigation are inconclusive since the device was not returned for analysis. Actions have been taken to prevent reoccurrence. Attempts were made to obtain complete patient information. Further information was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 7 ELITE
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8890144
MDR Text Key154188310
Report Number1627487-2019-09149
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/03/2020
Device Model Number3662
Device Lot Number6242834
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C

Patient Treatment Data
Date Received: 08/13/2019 Patient Sequence Number: 1
-
-