The device was not returned for analysis.
Since the device was not returned, we are unable to perform further root cause analysis.
All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.
If information is provided in the future, a supplemental report will be issued.
|
Medtronic receive report that after the uneventful navigation and deployment of the pipeline flex, when an attempt was made to raise microcatheter to collect the sleeve.
The delivery pusher was unable to be pulled in the middle.
The delivery pusher was pulled, and movement of the wire could not be found, so it was judged that the wire had broken.
However, the delivery pusher was almost retracted in microcatheter, and the delivery pusher was pulled out together with marksman.
The delivery pusher was removed from microcatheter outside the body, and the breakage was confirmed visually.
The break was in the proximal section of the pushwire.
There was no resistance felt during the use of the devices.
This event occurred during the treatment of an ica-cs that was not ruptured.
The max diameter was about 15mm with a neck width of 5mm.
The distal landing zone was 3.
4mm and the proximal was 4.
5mm.
The anatomy was moderate in tortuosity.
There was no patient injury.
There are no images.
The devices were prepared and used per the instructions for use (ifu).
|