Device evaluated by mfr.: returned product consisted of a nc emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 96.7cm from the hub and appears to have been kinked prior to separation.The inflation lumen and core with is kinked at the exit notch.Microscopic examination revealed no additional damages.There is blood present on the balloon folds.The balloon is still tightly folded.Although it was reported that the device was not used, the presence of blood on the balloon folds is indicative of handling beyond simply unpackaging the device, as was reported.Inspection of the remainder of the device presented no other damage or irregularities.
|
Reportable based on device analysis completed on 25-jul-2019.It was reported that shaft kink occurred.During preparation of a 3.50mm x 20mm nc emerge balloon catheter, it was noted that the device was kinked in the package.No patient complications were reported.However, returned device analysis revealed shaft detached/ separated.
|