Model Number DM3500 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Rn called tech services to report a patient who presented via merlin.Net device transmission with a 5-hour af episode.Upon review of the egm, tse discussed the episode was falsely triggered due to frequent pvcs the patient was experiencing.The patient will continue to be monitored, and rn and physician will discuss possible medication adjustment.
|
|
Event Description
|
It was reported that the patient presented remotely via merlin.Net transmission with atrial fibrillation (af) episode.Upon interrogation, the af episode was falsely triggered due to premature ventricular contractions (pvcs).Possible medication adjustment was discussed.The patient will be monitored.
|
|