Catalog Number SGC0301 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Exemption number (b)(4).The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report the steerable guide catheter leak.It was reported that during preparation of the steerable guide catheter (sgc) the hemostatic valve of the sgc, the hemostatic valve failed to hold column, three times.The device was not used, there was no patient involvement.A new sgc was used to complete the procedure.There was no clinically significant delay to the intended procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no lot specific quality issue from this lot.All available information was investigated and a definitive cause for the reported steerable guide catheter (sgc) leak (loss of fluid column during sgc preparation) could not be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|