Model Number 25-40-00
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/22/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the gas blender system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device.The service representative was not able to confirm the reported issue.As precaution the connecting cable will be replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a gas blender system displayed an error message on the panel as well on the module stating a defective module or sensor and air flow alarm during procedure.After swapping the device during procedure, no alarms were given and procedure could be finished with no further event.During testing after the procedure, the swapped device showed another error as well as the message stating a defective module or sensor.The allegation was made, that the connecting cable is defective.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
barbara galasso
14401 w. 65th way
arvada, CO 
MDR Report Key8909789
MDR Text Key175854296
Report Number9611109-2019-00639
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 08/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-