Device evaluated by the manufacturer.The hypotube, collar, distal shaft and tip was microscopically and visually inspected.Visual inspection revealed kinks to the hypotube 30.7 cm and 33.4 cm from the handle.The collar is separated, and the ptfe is still holding the two ends together.There are multiple flat spots along the distal shaft and the tip is flattened.Microscopic inspection revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found damages along the device that could have contributed to the difficulty to advance another device through the catheter.
|