• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE Back to Search Results
Catalog Number 03P84-25
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Abdominal Pain (1685)
Event Date 03/28/2019
Event Type  malfunction  
Manufacturer Narrative
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2019, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded a suspected discrepant creatinine results on a (b)(6) year old female patient with lower abdominal pain.There was no additional patient information available at the time of this report.Return product is not available for investigation.Method: i-stat, date: 03/28/2019, tested: 15:10, creatinine: 3.9.Er, 03/28/2019, n/a, 0.066.Collection time: unk.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 09/13/2019.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Ad, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CREA CARTRIDGE
Type of Device
CREA CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8915102
MDR Text Key195931770
Report Number2245578-2019-00198
Device Sequence Number1
Product Code CGL
UDI-Device Identifier10054749000125
UDI-Public10054749000125
Combination Product (y/n)N
PMA/PMN Number
K973292
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2019
Device Catalogue Number03P84-25
Device Lot NumberA18337A
Was Device Available for Evaluation? No
Date Manufacturer Received09/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age71 YR
-
-