Device evaluated by mfr.: returned product consisted of a guidezilla ii guide extension catheter.The hypotube, collar, distal shaft and tip was microscopically and visually inspected.Visual inspection revealed a kink to the hypotube 32.4cm from the handle.The tip is damaged so that it is no longer circular.Microscopic inspection revealed no additional damages.The device was x-rayed and there does not appear to be any abnormalities with the inside diameter of the shaft.The device was functionally tested by inserting a 0.014 guidewire and balloon through the inside diameter of the shaft.No resistance was felt.Inspection of the remainder of the device presented no other damage or irregularities.
|
It was reported that a device irregularity led to balloons tearing.The 90% stenosed target lesion was located on the moderately tortuous and severely calcified proximal and mid left anterior descending artery.A 6f guidezilla ii was selected for use.During the procedure, a guidezilla device was advanced to the lesion area of proximal lad.Two non-bsc balloons were advanced through the guidezilla ii to the lesion and both ruptured at 2-3 atmospheres (atm).When the balloons were checked outside the patient's body, they were observed to have longitudinal tears.It was suspected that there was a damage or irregularity in the guide segment / distal shaft of the device that make the balloons tear longitudinally.The devices were completely and simply pulled out from the patient's body.The procedure was completed with the original guidezilla ii device.There were no patient complications reported.Patient was in good condition post procedure.
|