• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problems Failure to Advance (2524); Material Protrusion/Extrusion (2979); Activation Problem (4042)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation, as the device was discarded after the event occurred.A lot history review (lhr) of redp3137 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (redp3137) have been reported from the same facility.
 
Event Description
It was reported "accucath catheter that we couldn¿t advance the wire to advance¿kept buckling inside the barrel and then wouldn¿t deploy the safety retraction.When i pulled the catheter off the needle before putting it in the sharps container, the thing deployed and the wire looped and was partially sticking out of the barrel as was part of the needle." it was reported this occurred with two accucaths.This report addresses the first device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUCATH INTRAVASCULAR CATHETER 20 GX2.25
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
kayla olsen
605 n. 5600 w.
salt lake city, UT 84116
8015225010
MDR Report Key8921485
MDR Text Key155354719
Report Number3006260740-2019-02455
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00801741110948
UDI-Public(01)00801741110948
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberAC0202250
Device Lot NumberREDP3137
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received08/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-