The reported firstpass suture passer device, intended for use in treatment, was not returned for evaluation.A relationship between the product and reported incident cannot be established as the product was not returned.Without the reported product a visual and functional evaluation cannot be performed and customer¿s complaint cannot be confirmed.From the information provided, ¿during a rotator cuff repair surgery the top jaw broke.¿ an exact root cause cannot be determined without evaluation of the device; however, factors unrelated to the manufacture or design of the device that could have contributed to the reported event include: (1) incorrect suture trap loading.(2) excessive force.(3) tissue thickness.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|